Ultrasound imaging

A linear probe sits on the phantom. Switch on the scanner, apply gel, choose the frequency and gain and measure the echo time with the time cursor in A-mode.

The screen shows the B-mode image (a cyst, a stone with its shadow, wire markers every 2 cm), the A-mode trace and the Doppler spectrum of flow in a vessel; the “Pulse-echo” view shows the pulse reflecting off layers.

In this lab

  • d = c·t/2, c ≈ 1540 m/s
  • R = ((Z₂ − Z₁)/(Z₂ + Z₁))²
  • λ = c/f
  • Δf = 2f₀·v·cosθ / c

What you learn

We study the pulse-echo method: find depth and the speed of sound in tissue from the echo time, see how frequency changes the image and measure blood flow speed with Doppler.

Laws and formulas

d = c·t / 2 — Echo depth: the pulse goes there and back.

R = ((Z₂ − Z₁)/(Z₂ + Z₁))² — Fraction reflected at a boundary, Z = ρc.

λ = c / f — The wavelength sets the resolution.

Δf = 2f₀·v·cosθ / c — The Doppler shift.

Equipment

  • Linear probe — a row of piezo crystals; 2–7.5 MHz.
  • Phantom — tissue-like material with fat, muscle, liver, a cyst, a stone, a vessel and wire markers every 2 cm.
  • Scanner screen — B-mode, A-mode and the Doppler spectrum.

Common mistakes

  • Forgetting the gel: air reflects all the energy.
  • Writing d = c·t: the pulse travels the path twice, so divide by 2.
  • A Doppler angle near 90°: cosθ ≈ 0 and the speed error explodes.

In real life

  • Pregnancies are followed with ultrasound because it uses no ionising radiation.
  • Echocardiography shows the heart valves and blood flow.
  • Bats and dolphins also “see” by echo.