Ultrasound imaging
A linear probe sits on the phantom. Switch on the scanner, apply gel, choose the frequency and gain and measure the echo time with the time cursor in A-mode.
The screen shows the B-mode image (a cyst, a stone with its shadow, wire markers every 2 cm), the A-mode trace and the Doppler spectrum of flow in a vessel; the “Pulse-echo” view shows the pulse reflecting off layers.
In this lab
- d = c·t/2, c ≈ 1540 m/s
- R = ((Z₂ − Z₁)/(Z₂ + Z₁))²
- λ = c/f
- Δf = 2f₀·v·cosθ / c
What you learn
We study the pulse-echo method: find depth and the speed of sound in tissue from the echo time, see how frequency changes the image and measure blood flow speed with Doppler.
Laws and formulas
d = c·t / 2 — Echo depth: the pulse goes there and back.
R = ((Z₂ − Z₁)/(Z₂ + Z₁))² — Fraction reflected at a boundary, Z = ρc.
λ = c / f — The wavelength sets the resolution.
Δf = 2f₀·v·cosθ / c — The Doppler shift.
Equipment
- Linear probe — a row of piezo crystals; 2–7.5 MHz.
- Phantom — tissue-like material with fat, muscle, liver, a cyst, a stone, a vessel and wire markers every 2 cm.
- Scanner screen — B-mode, A-mode and the Doppler spectrum.
Common mistakes
- Forgetting the gel: air reflects all the energy.
- Writing d = c·t: the pulse travels the path twice, so divide by 2.
- A Doppler angle near 90°: cosθ ≈ 0 and the speed error explodes.
In real life
- Pregnancies are followed with ultrasound because it uses no ionising radiation.
- Echocardiography shows the heart valves and blood flow.
- Bats and dolphins also “see” by echo.