Pulse oximeter
A sensor with red (660 nm) and infrared (940 nm) light is clipped to the patient's finger. Choose a patient, switch on the monitor and measure AC/DC at both beams.
The “Signal” view shows a finger cross-section and two photoplethysmograms marked with AC and DC; the “Spectrum” view shows the Hb, HbO₂ and COHb absorption curves.
In this lab
- I = I₀·e^(−ε·c·l)
- R = (AC/DC)₆₆₀ / (AC/DC)₉₄₀
- SpO₂ ≈ 110 − 25·R
- normal 95–100 %
What you learn
We study how a pulse oximeter works: find R from photoplethysmograms at two wavelengths, build a calibration line and determine SpO₂.
Laws and formulas
I = I₀·e^(−ε·c·l) — The Beer–Lambert law.
R = (AC/DC)₆₆₀ / (AC/DC)₉₄₀ — The ratio of ratios.
SpO₂ ≈ 110 − 25·R — An approximate calibration line.
PI = (AC/DC)₉₄₀ · 100 % — The perfusion index.
Equipment
- Finger sensor — red (660 nm) and infrared (940 nm) LEDs with a photodiode opposite.
- Monitor — SpO₂, pulse rate (PR), perfusion index (PI) and the photoplethysmogram.
Common mistakes
- A cold or moving finger gives a weak signal and an unreliable reading.
- Trusting SpO₂ in carbon monoxide poisoning: it reads too high.
- Dark nail polish blocks the light.
In real life
- Smartwatches use the same method with reflected light.
- Pilots and climbers watch their SpO₂ at altitude.