Tissue impedance

Place a piece of liver between silver electrodes, switch on the generator and measure |Z| and φ while changing the frequency from 100 Hz to 3 MHz.

The “Oscilloscope” view shows the U and I traces and |Z|(f) curves; the “Circuit” view shows the equivalent circuit, the Cole arc and the K values.

In this lab

  • |Z| = √(R² + X²)
  • X_C = 1/(2πfC)
  • tan φ = X/R
  • K = |Z(10 kHz)| / |Z(1 MHz)|

What you learn

We measure tissue impedance at different frequencies, observe the dispersion and judge viability with the polarisation coefficient.

Laws and formulas

|Z| = √(R² + X²) — Total impedance.

X_C = 1/(2πfC) — Capacitive reactance.

tg φ = X/R — Phase shift.

K = Z(10 kHz)/Z(1 MHz) — Tarusov’s coefficient.

Equipment

  • Measuring cell — a piece of liver between silver plate electrodes.
  • Generator and oscilloscope — 100 Hz to 3 MHz, U = 1 V; the U and I traces.

Common mistakes

  • Measuring with DC polarises the electrodes.
  • Overheating the sample damages the tissue.

In real life

  • Bioimpedance scales estimate body fat and water.
  • Rheography gauges organ blood filling from impedance changes.