Photocolorimetry
On the photoelectric colorimeter choose a filter and cuvette, zero on the blank, then measure KMnO₄ or haemoglobin standards and the unknown sample.
The “Spectrum” view shows the absorption curve and filters; the “Cuvette” view shows the light weakening, T, D and the D(c) graph.
In this lab
- T = I/I₀
- D = lg(I₀/I) = ε·c·l
- Hb: 540 nm, blood 1:251
- most accurate at D ≈ 0.1–1.0
What you learn
With a photoelectric colorimeter we test the Beer–Lambert law, choose the best filter and use a calibration graph to find an unknown concentration, including blood haemoglobin.
Laws and formulas
T = I/I₀ — Transmittance.
D = lg(I₀/I) = −lg T — Optical density.
D = ε·c·l — The Beer–Lambert law.
Equipment
- Photoelectric colorimeter — a lamp, 7 filters (400–750 nm), a cuvette compartment and a photocell.
- Cuvettes — path lengths 0.5, 1, 2 and 3 cm.
- Standard solutions — KMnO₄ 0.1–0.4 mmol/l; Hb 60–180 g/l.
Common mistakes
- Not re-zeroing after changing the filter.
- Measuring at very large or very small D.
- A fingerprint or bubble on the cuvette wall.
In real life
- Clinical chemistry analysers rest on the same law.
- The pulse oximeter uses the Beer–Lambert law too.