Photocolorimetry

On the photoelectric colorimeter choose a filter and cuvette, zero on the blank, then measure KMnO₄ or haemoglobin standards and the unknown sample.

The “Spectrum” view shows the absorption curve and filters; the “Cuvette” view shows the light weakening, T, D and the D(c) graph.

In this lab

  • T = I/I₀
  • D = lg(I₀/I) = ε·c·l
  • Hb: 540 nm, blood 1:251
  • most accurate at D ≈ 0.1–1.0

What you learn

With a photoelectric colorimeter we test the Beer–Lambert law, choose the best filter and use a calibration graph to find an unknown concentration, including blood haemoglobin.

Laws and formulas

T = I/I₀ — Transmittance.

D = lg(I₀/I) = −lg T — Optical density.

D = ε·c·l — The Beer–Lambert law.

Equipment

  • Photoelectric colorimeter — a lamp, 7 filters (400–750 nm), a cuvette compartment and a photocell.
  • Cuvettes — path lengths 0.5, 1, 2 and 3 cm.
  • Standard solutions — KMnO₄ 0.1–0.4 mmol/l; Hb 60–180 g/l.

Common mistakes

  • Not re-zeroing after changing the filter.
  • Measuring at very large or very small D.
  • A fingerprint or bubble on the cuvette wall.

In real life

  • Clinical chemistry analysers rest on the same law.
  • The pulse oximeter uses the Beer–Lambert law too.