Electrocardiography

Electrodes are clipped to the mannequin’s limbs. Choose a patient, switch on the electrocardiograph, hold the recording and measure the strip with the cursors.

The “Einthoven” view shows the heart vector and its projections on the three leads; the “ECG strip” shows P, QRS and T at 25 mm/s and 10 mm/mV.

In this lab

  • II = I + III: Einthoven’s law
  • HR = 60 / RR
  • tan α = (I + 2·III) / (√3·I)
  • normal axis −30° to +90°

What you learn

We study the heart as an electric dipole: record an ECG in the standard leads, check Einthoven’s law and find the heart rate and electrical axis.

Laws and formulas

I = φ_LA − φ_RA, II = φ_LL − φ_RA, III = φ_LL − φ_LA — The standard leads.

II = I + III — Einthoven’s law.

HR = 60 / RR — Heart rate with RR in seconds.

tg α = (I + 2·III) / (√3·I) — The electrical-axis angle.

Equipment

  • Electrodes — RA red, LA yellow, LL green, RL black (ground).
  • Electrocardiograph — paper 25 or 50 mm/s, gain 5, 10 or 20 mm/mV.

Common mistakes

  • Swapping the arm electrodes inverts lead I.
  • Forgetting the paper speed: at 50 mm/s, 1 mm = 0.02 s.

In real life

  • Heart monitors and smartwatches measure the same potential differences.
  • A Holter monitor records the ECG for a whole day.