Electrocardiography
Electrodes are clipped to the mannequin’s limbs. Choose a patient, switch on the electrocardiograph, hold the recording and measure the strip with the cursors.
The “Einthoven” view shows the heart vector and its projections on the three leads; the “ECG strip” shows P, QRS and T at 25 mm/s and 10 mm/mV.
In this lab
- II = I + III: Einthoven’s law
- HR = 60 / RR
- tan α = (I + 2·III) / (√3·I)
- normal axis −30° to +90°
What you learn
We study the heart as an electric dipole: record an ECG in the standard leads, check Einthoven’s law and find the heart rate and electrical axis.
Laws and formulas
I = φ_LA − φ_RA, II = φ_LL − φ_RA, III = φ_LL − φ_LA — The standard leads.
II = I + III — Einthoven’s law.
HR = 60 / RR — Heart rate with RR in seconds.
tg α = (I + 2·III) / (√3·I) — The electrical-axis angle.
Equipment
- Electrodes — RA red, LA yellow, LL green, RL black (ground).
- Electrocardiograph — paper 25 or 50 mm/s, gain 5, 10 or 20 mm/mV.
Common mistakes
- Swapping the arm electrodes inverts lead I.
- Forgetting the paper speed: at 50 mm/s, 1 mm = 0.02 s.
In real life
- Heart monitors and smartwatches measure the same potential differences.
- A Holter monitor records the ECG for a whole day.