Pulse oximeter

A sensor with red (660 nm) and infrared (940 nm) light is clipped to the patient's finger. Choose a patient, switch on the monitor and measure AC/DC at both beams.

The “Signal” view shows a finger cross-section and two photoplethysmograms marked with AC and DC; the “Spectrum” view shows the Hb, HbO₂ and COHb absorption curves.

In this lab

  • I = I₀·e^(−ε·c·l)
  • R = (AC/DC)₆₆₀ / (AC/DC)₉₄₀
  • SpO₂ ≈ 110 − 25·R
  • normal 95–100 %

What you learn

We study how a pulse oximeter works: find R from photoplethysmograms at two wavelengths, build a calibration line and determine SpO₂.

Laws and formulas

I = I₀·e^(−ε·c·l) — The Beer–Lambert law.

R = (AC/DC)₆₆₀ / (AC/DC)₉₄₀ — The ratio of ratios.

SpO₂ ≈ 110 − 25·R — An approximate calibration line.

PI = (AC/DC)₉₄₀ · 100 % — The perfusion index.

Equipment

  • Finger sensor — red (660 nm) and infrared (940 nm) LEDs with a photodiode opposite.
  • Monitor — SpO₂, pulse rate (PR), perfusion index (PI) and the photoplethysmogram.

Common mistakes

  • A cold or moving finger gives a weak signal and an unreliable reading.
  • Trusting SpO₂ in carbon monoxide poisoning: it reads too high.
  • Dark nail polish blocks the light.

In real life

  • Smartwatches use the same method with reflected light.
  • Pilots and climbers watch their SpO₂ at altitude.