Mass spectrometer
The source ionises the gas and accelerates the ions; an electromagnet bends them in a vacuum chamber. A detector slides along the scale and measures the current where each beam lands.
The “From above” view shows the isotope beams fanning out; “Spectrum” shows the detector current and mass spectrum, isotope shares and average atomic mass.
In this lab
- qU = mv²/2: acceleration
- qvB = mv²/r: the Lorentz force
- r = √(2mU/q)/B: the radius
- isotopes and average atomic mass
What you learn
We accelerate ions with an electric field and bend them with a magnetic one, see how the radius depends on mass, voltage and field, measure the proton’s charge-to-mass ratio and identify the isotopes of an unknown gas.
Laws and formulas
qU = mv²/2 — The electric field’s work becomes the ion’s kinetic energy.
qvB = mv²/r — The Lorentz force provides the centripetal force.
r = √(2mU/q) / B — The radius goes as the square root of the mass.
m/q = B²x²/(8U) — x = 2r is the distance to the landing point.
Equipment
- Ion source — ionises the gas and accelerates the ions at 500–3000 V.
- Electromagnet — a uniform field of 0.05–0.3 T.
- Vacuum chamber — so the ions do not hit air molecules.
- Detector — slides along the scale; an electrometer measures the ion current.
Common mistakes
- Measuring the radius instead of x = 2r and getting a mass 4 times too small.
- Mistaking the H₂⁺ peak for the proton.
- Reading on the edge of a peak, which gives too little current.
In real life
- Radiocarbon dating gives the age of ancient finds.
- Detecting doping in sport and poisons in food.
- Isotopes are separated when enriching uranium for nuclear power.